Microscope Alignment

Menu: Extensions > QP Scope > Microscope Alignment Back to README | All Tools | All Workflows

Purpose

Create or update the coordinate transformation between QuPath image coordinates and physical microscope stage positions. This alignment is required for accurate stage positioning when acquiring from existing images. Use this tool the first time you set up a slide/scanner combination, after hardware changes affecting positioning, or when acquired images do not align with annotations.

Microscope Alignment source selection dialog

Prerequisites

  • Macro/overview image loaded in QuPath with pixel size calibration
  • Microscope server connected
  • Stage can move to known positions
  • Image must have visible features suitable for point matching

Options

Macro Image Selection

Option Type Description
Image ComboBox Select the macro/overview image to use for alignment (if multiple images are in the project)

The selected image must have pixel size calibration and should cover the area of interest.

Point Marking

Column Type Description
# Label Point number
Image X Label X coordinate in QuPath pixels
Image Y Label Y coordinate in QuPath pixels
Stage X Label X coordinate in micrometers
Stage Y Label Y coordinate in micrometers
Delete Button Remove this point pair

Mark at least 3 points. More points improve accuracy.

Transform Quality Metrics

Metric Good Value Description
Mean Error < 50 um Average positioning error across all points
Max Error < 100 um Worst-case error for any single point
R-squared > 0.99 Overall fit quality (1.0 = perfect)

Save Transform

Field Type Description
Transform Name TextField Descriptive name for this transform
Scanner ComboBox Scanner type this applies to (if applicable)
Notes TextArea Optional notes about conditions or setup

Workflow

Step 1: Macro Image Selection

If multiple images exist in the project, select the macro/overview image to use for alignment. The image must have visible features that you can also locate under the microscope.

The alignment workflow automatically loads the flipped version of the macro image (matching the Live Viewer orientation) before point selection begins. This ensures that image coordinates correspond correctly to what you see through the camera.

Step 2: Point Marking

For each calibration point:

  1. In QuPath: Click on a recognizable feature in the macro image
  2. Move Stage: Navigate the microscope stage to the same physical location
  3. Record Point: Click "Record Point" to capture both coordinate pairs
  4. Repeat: Mark at least 3 points (more points = better accuracy)

Point distribution guidelines:

  • Spread points across the entire image
  • Cover corners and center if possible
  • Do not cluster points in one area
  • Use easily identifiable features (tissue edges, landmarks)

Step 3: Transform Calculation

After marking points, click Calculate Transform. The system computes the best-fit affine transformation and displays error statistics. Review the quality metrics to assess accuracy.

Step 4: Refinement (Manual or Automatic)

Refine Alignment dialog

After the initial transform is calculated, single-tile refinement improves accuracy:

  1. Select a tile in the image
  2. The stage moves to the predicted position
  3. Refine using one of two methods:

Auto-Align (SIFT) -- recommended when tissue features are visible:

  • Click the Auto-Align (SIFT) button in the refinement dialog
  • The system extracts a region from the WSI around the selected tile (with 160um search margin)
  • A microscope image is snapped and matched against the WSI region using SIFT feature detection
  • Pixel size differences between the WSI and microscope are automatically handled (both images are rescaled to the lower resolution)
  • Optical flip correction is applied based on image metadata
  • The stage moves to correct any offset found by the matching
  • The status shows the offset and number of matched features
  • Verify the live view matches the tile, then click "Save Refined Position"

Manual alignment -- fallback when SIFT cannot find enough features:

  • Use the stage controls or Live Viewer to manually adjust the position
  • Match the live microscope view to the selected tile in QuPath
  • Click "Save Refined Position"

SIFT auto-alignment works best on tissue with visible structural features. It may fail on blank areas, very uniform tissue, or regions with repetitive patterns. In those cases, use manual alignment.

Trust SIFT mode (advanced): Enable via the trustSiftAlignment preference (see Preferences). When enabled, the refinement step runs SIFT automatically without showing the manual dialog. If the confidence (inlier ratio) exceeds the configurable threshold (default 50%), the position is auto-accepted and the workflow continues unattended. Falls back to the manual dialog if SIFT fails or confidence is too low.

Step 5: Validation

Test the calculated transform:

  1. Select a validation point (not used in the calculation)
  2. Click Go To Point to move the stage
  3. Verify the stage arrives at the expected physical location
  4. Check visual alignment through the microscope eyepiece or live viewer

If validation fails:

  • Add more calibration points
  • Check for systematic errors (flip/invert settings)
  • Ensure points were accurately marked in both coordinate systems

Step 5: Quality Summary and Save

After calculating the transform, a quality summary dialog is displayed with pass/warn/fail indicators for:

  • Scale -- Whether the computed pixel-to-stage scale matches the expected value
  • Residuals -- Whether the point-fitting residual errors are within acceptable limits
  • Translation stability -- Whether the translation component is consistent across point subsets

If any indicator shows a warning or failure, you must acknowledge the issue before the alignment is saved. This prevents silently saving a poor-quality transform.

Give the transform a descriptive name and save it. Saved transforms are stored in the configuration folder and can be selected in the Existing Image Acquisition workflow. Multiple transforms can exist for different conditions.

Output

  • A saved coordinate transformation file linking QuPath pixel coordinates to stage positions
  • Transform can be reused across sessions until hardware or slide position changes
  • Quality metrics for assessing alignment accuracy

Understanding Flip and Invert Settings

These are two independent concepts that affect coordinate mapping:

Image Flipping (Optical Property):

  • Corrects for optical inversions in the microscope light path
  • Affects how annotations appear relative to stage coordinates
  • Set in Preferences: "Flip macro image X/Y"

Stage Inversion (Coordinate Direction):

  • Corrects for stage coordinate system conventions
  • Affects which direction stage movement commands go
  • Set in Preferences: "Inverted X/Y stage"

Typical Settings:

Configuration Flip X Flip Y Invert X Invert Y
Standard upright OFF OFF OFF ON
Inverted microscope OFF ON OFF ON
Flipped slide scanner ON OFF OFF ON

Tips & Troubleshooting

Issue Cause Solution
Large mean error Points inaccurately marked Re-mark with more care and precision
Stage moves wrong direction Invert settings wrong Toggle Invert X or Y in Preferences
Image appears mirrored Flip settings wrong Toggle Flip X or Y in Preferences
Points do not form a pattern Incorrect point pairs Verify each QuPath point matches its stage position
Validation point is off Not enough calibration points Add more points spread across the image